Introduction:

This procedure is used to install a Hotfix for the Dr. Oligo Control Software

Applies to:

This procedure applies to the following:

Part NumberDescription
403-107733Dr. Oligo 192c
403-108068Dr. Oligo 192XLc
403-108163Dr. Oligo 768XLc
OtherOther Dr. Oligo Plate Instruments running software version 6.0 or greater

Prerequisites:

  1. A link for the software updater from support.

Procedure:

When there is a new version of software to update to for the Dr. Oligo Control follow this procedure.

  1. Download the new version of software:
    • Please contact support for the link to the latest software needed for your instrument.
    • Once you have downloaded the Updater, copy it to the Desktop of the PC that is running the Dr. Oligo Control software.
  2. Backup Existing Data Files:
    • To make sure that you do not lose any existing data please perform the following backup before running the new software updater. These instructions will tell you how to make a backup of your existing data.
    • Backup the "C:\ProgramData\Biolytic Lab Performance" Folder
      1. Open a windows explorer window (press the windows key on the keyboard and the “W” at the same time)
      2. In the left side of this window select the Windows (C:) drive
      3. In the right window right click. A popup menu will appear at your mouse pointer. 
        1. Put the mouse pointer on the “New” selection
        2. A fly out menu will appear and one of the selections is Folder, select Folder.
        3. The default new folder name is going to be: New folder. This name is ok to use.
      4. In the left-hand window under the Windows (C:) drive click on the folder named “ProgramData”
      5. In the right-hand window, you will see a folder named “Biolytic Lab Performance”
      6. Right click on the Biolytic Lab Performance folder and select “Copy” from the popup menu.
      7. In the left-hand window click the “Windows (C:)” drive
      8. The right-hand window will now list folders and one of them is the New folder that you created earlier. (step 3.3)
      9. Double click the “New folder” in the right-hand window.
      10. The right-hand window will not be blank (no files or folders)
      11. Right click in the right-hand window and select “Paste” from the popup menu.
      12. Wait for the copy to complete.
      13. Now a copy of your data files are all saved in the “New Folder”.
    • Backup the "C:\Program Files (x86)\Dr. Oligo" Folder
      1. In the left-hand window under the Windows (C:) drive click on the folder named “Program Files (x86)”
      2. In the right-hand window, you will see a folder named “Dr. Oligo”
      3. Right click on the Biolytic Lab Performance folder and select “Copy” from the popup menu.
      4. In the left-hand window click the “Windows (C:)” drive
      5. The right-hand window will now list folders and one of them is the New folder that you created earlier. (step 3.3 in the previous section)
      6. Double click the “New folder” in the right-hand window.
      7. The right-hand window will not be blank (no files or folders)
      8. Right click in the right-hand window and select “Paste” from the popup menu.
      9. Wait for the copy to complete.
      10. Now a copy of your data files are all saved in the “New Folder”.
  3. Performing the Software Update:
    1. Now, double click the updater file that you copied to the desktop on the PC that is controlling the instrument.
    2. This will run the updater.
    3. Follow the instructions and select the default values.
    4. Run the new software. The startup screen should indicate that it is the new version.
    5. Check that your protocols, macros, synthesis files, logs, etc. are all present in the updated software.
  4. Save the backup folders for 30 days after you install the update and make sure that the protocols all work as they did before.

Troubleshooting:

  1. The updater link is expired
    • Contact support for an updated link
  2. The software license is invalid
    • Contact support and include the following information
      • The software version you are updating from
      • The hotfix version you are updating to
      • A screenshot of the login screen
      • The Instrument Serial Number
      • A copy of the following folders:
        • "C:\Program Files (x86)\Dr. Oligo"
        • "C:\ProgramData\Biolytic Lab Performance\DrOligo"
 
 
 
 
Related Products Dr. Oligo 192c Oligo Synthesizer, Dr. Oligo 192XLc Oligo Synthesizer, Dr. Oligo 768XLc Oligo Synthesizer