Introduction:
This procedure is used to install a Hotfix for the Dr. Oligo Control Software
Applies to:
This procedure applies to the following:
Part Number | Description |
403-107733 | Dr. Oligo 192c |
403-108068 | Dr. Oligo 192XLc |
403-108163 | Dr. Oligo 768XLc |
Other | Other Dr. Oligo Plate Instruments running software version 6.0 or greater |
Prerequisites:
- A link for the software updater from support.
Procedure:
When there is a new version of software to update to for the Dr. Oligo Control follow this procedure.
- Download the new version of software:
- Please contact support for the link to the latest software needed for your instrument.
- Once you have downloaded the Updater, copy it to the Desktop of the PC that is running the Dr. Oligo Control software.
- Backup Existing Data Files:
- To make sure that you do not lose any existing data please perform the following backup before running the new software updater. These instructions will tell you how to make a backup of your existing data.
- Backup the "C:\ProgramData\Biolytic Lab Performance" Folder
- Open a windows explorer window (press the windows key on the keyboard and the “W” at the same time)
- In the left side of this window select the Windows (C:) drive
- In the right window right click. A popup menu will appear at your mouse pointer.
- Put the mouse pointer on the “New” selection
- A fly out menu will appear and one of the selections is Folder, select Folder.
- The default new folder name is going to be: New folder. This name is ok to use.
- In the left-hand window under the Windows (C:) drive click on the folder named “ProgramData”
- In the right-hand window, you will see a folder named “Biolytic Lab Performance”
- Right click on the Biolytic Lab Performance folder and select “Copy” from the popup menu.
- In the left-hand window click the “Windows (C:)” drive
- The right-hand window will now list folders and one of them is the New folder that you created earlier. (step 3.3)
- Double click the “New folder” in the right-hand window.
- The right-hand window will not be blank (no files or folders)
- Right click in the right-hand window and select “Paste” from the popup menu.
- Wait for the copy to complete.
- Now a copy of your data files are all saved in the “New Folder”.
- Backup the "C:\Program Files (x86)\Dr. Oligo" Folder
- In the left-hand window under the Windows (C:) drive click on the folder named “Program Files (x86)”
- In the right-hand window, you will see a folder named “Dr. Oligo”
- Right click on the Biolytic Lab Performance folder and select “Copy” from the popup menu.
- In the left-hand window click the “Windows (C:)” drive
- The right-hand window will now list folders and one of them is the New folder that you created earlier. (step 3.3 in the previous section)
- Double click the “New folder” in the right-hand window.
- The right-hand window will not be blank (no files or folders)
- Right click in the right-hand window and select “Paste” from the popup menu.
- Wait for the copy to complete.
- Now a copy of your data files are all saved in the “New Folder”.
- Performing the Software Update:
- Now, double click the updater file that you copied to the desktop on the PC that is controlling the instrument.
- This will run the updater.
- Follow the instructions and select the default values.
- Run the new software. The startup screen should indicate that it is the new version.
- Check that your protocols, macros, synthesis files, logs, etc. are all present in the updated software.
- Save the backup folders for 30 days after you install the update and make sure that the protocols all work as they did before.
Troubleshooting:
- The updater link is expired
- Contact support for an updated link
- The software license is invalid
- Contact support and include the following information
- The software version you are updating from
- The hotfix version you are updating to
- A screenshot of the login screen
- The Instrument Serial Number
- A copy of the following folders:
- "C:\Program Files (x86)\Dr. Oligo"
- "C:\ProgramData\Biolytic Lab Performance\DrOligo"
- Contact support and include the following information