Views:

Introduction:

An article showing the location of the main gas valve on the Dr. Oligo 192c and how it is connected to the Biolytic board.

Applies to:

This document applies to the following.

Part Number Description
403-107733 Dr. Oligo 192-C Plate Oligo Synthesizer Instrument
615-107775 Main Gas Valve Pneumatic Assembly
181-108033 Biolytic Board

Procedure:

  1. Looking at the back of the instrument, the Left Panel is on the right-hand side of the instrument. There you will find several important valves, including the Main Gas Valve.

  2. The Main Gas Valve will be near the bottom of the Left Panel, close to the regulator. On the valve there should be a wire connector. Make sure that the connector is securely attached to the valve and that the wires are secure in the connector.


     
  3. The wire from the Main Gas Valve then goes to the Biolytic Board 1.


     
  4. The Main Gas Valve wiring should connect to Channel 1 which is in the back of the row on the right side of the board. Make sure the connection is secure and that the wires are securely in the connector.

 
 
 
 
 